Process Calibration Tools Catalog EN

Suggested test tools Gas custody transfer flow computer calibration Gas custody transfer flow computers that calculate flow in pipelines by measuring the differential pressure across a flow restriction, such as an orifice plate or other differential pressure flow device, require special calibration to perform at optimum accuracy. Gas flow computers make three primary measurements to calculate flow: volumetric flow (difference in pressure across the orifice plate), static pressure in a pipeline and gas temperature. A calculation is performed using this data to determine the actual mass and volume of the gas flowing through the pipeline. These calibrations can be made with three separate calibrators, a low pressure, high pressure and a temperature calibrator or use a multifunction calibration tool designed for this specific task. An example of a calibrator purposed for this task is the Fluke 721 or 721Ex. It has two built- in pressure ranges and the ability to measure temperature. The most popular configuration is 16 psi/1 bar on the low pressure (P1) sensor side and 1500/100 bar or 3000 psi/200 bar on the high pressure (P2) sensor side. It measures temperature using a precision RTD accessory and can display all three measurements at once if desired. Fluke 754 Documenting Process Calibrator-HART See pg 5 Fluke 700G Precision Pressure Gauge Calibrator See pg 13 Fluke 721 Precision Dual Range Pressure Calibrator See pg 12 Fluke 750P Series Pressure Modules See pg 12 34 Pressure Applications