712/714 Temperature Calibrators
80PK-24 (714)
See page 136
The Fluke 710 Series Process Calibrators
offer clear new choices in single-function
calibrators. Whatever you want to
measure - temperature, pressure or basic
electrical parameters - you’ll find one of
these easy-to-carry handheld tools gives
you exactly the functions you need. They
combine the rugged, ready-for-action
package of the proven Fluke 80 Series
DMMs with the easy-to-use pushbutton
operation of the multifunction Fluke 740
Series Documenting Process Calibrators.
These calibrators are EMI tolerant,
dust and splash resistant, and have a
removable battery door for quick battery
changes.
712 RTD Calibrator
• Measure temperature from RTD output
• Simulate RTD output
• Rosemount pulsed RTD transmitter
compatible
• Operates with seven types of RTDs
• Measure additional RTDs using Ohms
measurement function
• Simulate additional RTDs using Ohms
source function
• °F or °C selectable
• Four shrouded banana jacks
714 Thermocouple Calibrator
• Measure temperature fromTC output
• Simulate TC output
• Operable with nine types of
thermocouples
• Calibrate linear TC transmitter with
mV source function
• Selectable °F or °C
• Thermocouple mini-jack termination
• Available as accessories; Fluke 700 TC1
and TC2 Thermocouple Mini-plug Kits
Fluke 714
Included Accessories
Yellow Holster (H80M excl. TPAK), Test Leads
and AC175 Alligator Clips (excluding model
714), single 9V alkaline battery and Instruction
Sheet (14 languages)
Ordering Information
Fluke 712
RTD Calibrator
Fluke 714
Thermocouple Calibrator
Recommended Accessories
TL220-1 (714)
See page 130
TL970
See page 129
C550
See page 138
C25
See page 138
Maximum voltage:
30 V
Operating temperature:
-10°C to 55°C
Safety:
CSA C22.2 No. 1010.1:1992
EMC: EN50082-1:1992 and EN55022:
1994 Class B
Size (HxWxD):
201mm x 98 mm x 52 mm
Weight:
0.6kg
Battery:
9V alkaline
Battery Life:
4 to 20 hours, typical, depending
on functions used
Three Year Warranty
Model
Function
Range
Resolution
Accuracy
Notes
Fluke 712 Measure/
simulate RTD
-200 to 800ºC
(Pt 100)
0.1ºC, 0.1ºF
0.33ºC, 0.6ºF
(Pt 100)
Pt; 100, 200, 500,
1000 (385);
Pt 100 (3926);
Pt 100 (3916) JIS;
Ni 120 (672)
Measure/
simulate
Resistance
15 Ω to 3200 Ω 0.1 Ω
0.1 Ω to 1 Ω
Fluke 714 Measure/
simulate
Thermocouple
-200 to 1800ºC,
depending on
type (K, -200 to
1370ºC)
0.1–C or ºF
(1ºC or ºF; BRS)
0.3ºC to 10 μV 9 TC types:
J K T E R S B per NIST
175 and ITS-90, L U per
DIN 43710 and IPTS-68
Measure/
simulate mV
-10 to 75 mv
0.01 mV
0.025% +
1 count
Specifications
Fluke 712
The clear choice