Best match |
---|
Fluke 355 |
Buy Now |
Best match |
|||
---|---|---|---|
Fluke 1630-2 FC |
|||
Buy Now | |||
Electrical specifications | |||
Maximum voltage to earth ground | 1000V | ||
Battery type | 4 x 1.5 V AA alkaline IEC/EN LR6 | ||
Battery life | More than 15 hours* | ||
Frequency range | 40 Hz to 1 kHz | ||
Ingress protection | IEC/EN 60529: IP30 with jaw closed | ||
LCD - Digital reading | 9999 counts | ||
LCD - Refresh rate | 4/s | ||
Operating temperature | -10 °C to +50 °C (14 °F to +122 °F) | ||
Storage temperature | -20 °C to +60 °C (-4 °F to +140 °F) | ||
Operating humidity | Non condensing (<10 °C) (< 50 °F)
≤90 % RH (at 10 °C to 30 °C [50 °F to 86 °F]) ≤75 % RH (at 30 °C to 40 °C [86 °F to 104 °F]) ≤45 % RH (at 40 °C to 50 °C [104 °F to 122 °F]) (Non condensing) |
||
Operating altitude | 2000 m (6561 ft.) | ||
Storage altitude | 12,000 m (39370 ft.) | ||
Reference temperature | 23 °C ±5 °C (73 °F ±9 °F) | ||
Temperature coefficient | 0.15 % X (specified accuracy)/ °C ( <18 °C or>28 °C ([ <64.4 °F or>82.4 °F]) | ||
Overload indication | OL | ||
Downloads | |||
Datasheet | Download | ||
The Fluke 1630 Earth Ground Clamp Meter is able to measure ground loop resistances—using the stakeless testing method. This test technique eliminates the dangerous, and time consuming activity, of disconnecting parallel grounds. You can also perform earth ground tests: inside buildings, on power pylons or anywhere you don’t have access to soil. With the stakeless testing method, no earth ground stakes are used at all. A known voltage is induced by one half of the clamp and the current is measured by the other half. The tester automatically determines the ground loop resistance at this grounding connection.