Fluke 773 |
Fluke 772 |
Fluke 771 |
||||
|
|
|
||||
|
|
|
|
||||
| Features | ||||||
| mA measure w/jaw | • | • | • | |||
| mA measure In circuit | • | • | ||||
| mA source | • | • | ||||
| mA sim | • | • | ||||
| Loop power 24 V | • | • | ||||
| DCV source 0-10 V | • | |||||
| DCV measure 0-30 V | • | |||||
| Scaled mA output to mA input | • | |||||
| mA in/out | • | |||||
| Specifications | ||||||
| mA measurement — measured by clamp:
Resolution/range: 0 mA to 20.99 mA / Accuracy : 0.2% +5 counts Resolution/range: 21.0 mA to 100 mA / Accuracy : 1% +5 counts |
• | • | • | |||
| mA measurement — measured in series with test jacks:
Resolution/range: 0 mA to 24.00 mA / Accuracy : 0.2% +2 counts |
• | • | ||||
| mA source — maximum mA drive: 24 mA into 1,000 ohms:
Resolution/range: 0 mA to 24.00 mA / Accuracy : 0.2% +2 counts |
• | • | ||||
| mA simulate — maximum voltage 50 V dc:
Resolution/range: 0 mA to 24.00 mA / Accuracy : 0.2% +2 counts |
• | • | ||||
| Voltage source — 2 mA maximum drive current:
Resolution/range: 0 V dc to 10.00 V dc / Accuracy : 0.2% +2 counts |
• | |||||
| Voltage measure:
Resolution/range: 0 V dc to 30.00 V dc / Accuracy : 0.2% +2 counts |
• | |||||
| Downloads | ||||||
| Datasheet | Download | Download | ||||
The Fluke 771, 772 and 773 mA Process Clamp Meters pay for themselves. These meters save time by measuring 4 to 20 mA signals (five) times faster than traditional DMMs without opera- tor intervention, offering loop calibration, voltage testing and eliminate time wasting activities.
Or use a Current Meter Module: