149
Current Clamps
80i-110s
i30
i30s
i310s
i410
i1010
Specifications AC/DC Models
* For specifications of 90i-610s see page 154
1 For DC only
2 Requires PM 9081 (see page 103)
3 Requires PM 9082 (see page 103)
4 115, 117 only
1)
Basic Accuracy: % reading + floorspec
80i-110s
i30
i30s
i310s
i410
i1010
Measurement type
Hall sensor
Hall sensor
Hall sensor
AC/DC
Hall sensor
Hall sensor
Nominal current range(s)
10 A, AC/DC
100 A, AC/DC
20 A AC RMS or DC 20 A AC RMS or DC 30/300 A AC RMS
or 45/450 A DC
400 A, AC/DC
600 A, AC
1000 A, DC
Continuous current range
0.1 A - 10 A AC/DC
1 A - 100 A AC/DC
30 A AC Peak
30 A AC Peak
100 mA - 300 A AC
RMS or 450 A DC
1 A - 400 A AC/DC
1 A - 600 A, AC
1 A - 1000 A, DC
Highest current
140 A - 2 kHz
30 A AC Peak
30 A AC Peak
300 A AC RMS or
450 A DC
400 A
1000 A
Lowest measurable current
0.1 A
50 mA
50 mA
100 mA
0.5 A
0.5 A
Basic accuracy
1)
3% + 50 mA (@ 10 A)
± 1% of reading ±2 mA ± 1% of reading ±2 mA ±1% of reading
3.5% + 0.5 A
2% +0.5 A
Useable frequency
DC - 100 kHz
DC to 20 kHz (-0.5dB)
DC to 100 kHz (-0.5dB) DC to 20 kHz
DC - 3 kHz
DC - 10 kHz
Zero error adjustment
●
manual adjust via
thumbwheel
manual adjust via
thumbwheel
Manual
●
●
Max. working voltage
600 V
300 V AC RMS
300 V AC RMS
300 V AC RMS or DC 600 V
600 V
Maximum conductor diameter
11.8 mm
19 mm
19 mm
19 mm
30 mm
2 x 25 mm
30 mm
2 x 25 mm
Output level(s)
100 mV/A
10 mV/A
100 mV/A
100 mV/A
10/1 mV/A
1 mV/A
1 mV/A
Battery, battery life
9 V, 55 h
30 hours typical
30 hours typical
30 hours
9 V, 60 h
9 V, 60 h
Output cable length (m)
1.6
1.5
2
2
1.6
1.6
Shrouded banana Plugs
●
n/a
●
●
BNC connector
●
n/a
●
●
BNC to banana adapter included
n/a
●
●
Safety
CAT II, 600 V
CAT III, 300 V
CAT III, 300 V
CAT III, 300 V
CAT III 300 V
CAT III, 600 V
CAT III, 600 V
Current Clamp Compatibility Chart
113/114/115/116/117
175/177/179
233
287/289
27II / 28II
3000FC
8845A/8846/8808A
77 IV
83V/87V/88V
43B
430 Series II
123/124/125
190 Series II
1577/1587
715/724/725
753/754
787/789
AC Models
i5s
●
●
i200
4
● ● ● ● ● ● ● ●
●
●
i200s
● ● ● ● ● ● ● ● ●
● ● ●
● ●
i400
● ● ● ● ● ● ● ●
●
●
i400s
2 2 2 2 2 2 2 2 2
● ● ● ●
2
2 2
i430 flexi-TF-II
●
i800
4
● ● ● ● ● ● ● ●
●
●
i1000s
2 2 2 2 2 2 2 2 2
●
● ●
2
2 2
i2000 flex (new version)
● ● ● ● ● ● ● ● ●
3 3 3 3
●
●
i3000s
● ● ● ● ● ● ● ● ● ● ● ● ● ●
● ●
i3000s flex
● ● ● ● ● ● ● ● ● ● ● ● ● ●
● ●
i6000s flex
● ● ● ● ● ● ● ● ●
●
●
● ●
AC/DC Models
i30
● ● ● ● ● ● ● ● ●
●
●
i30s
● ● ● ● ● ● ● ● ● ● ● ● ●
80i-110s
2 2 2 2 2 2 2 2 2
● ● ● ● ●
2 2
i310s
● ● ● ● ● ● ● ● ● ● ● ● ● ●
● ●
i410 / i410 kit
● ● ● ●
● ● ● ●
3 3
●
1
● ●
i1010 / i1010 kit
● ● ● ● ● ● ● ● ●
3 3
●
1
● ●
Other
90i-610s*
2 2 2 2 2 2 2 2 2
● ●
2
2 2
All accessories have a one year warranty
i410 Kit AC/DC Current Clamp
(400A) with soft case
i1010 Kit AC/DC Current Clamp
(1000A) with soft case
• Combination of current clamp with
carrying case
• Zipped soft case with moveable
divider
• Soft case is large enough to hold a
meter