61
Separate hard case
The compact Fluke PAT testers are supplied with a
hard carrying case that not only offers protection
during transit but also includes extra storage space for
accessories and other tools. They’re extremely light,
weighing approximately 3 kg (without case) and have
integral carrying handles for extra convenience.
Special PAT Kit
If you need a complete PAT tester solution,
a purpose made kit is available.
Fluke 6500-2/UK Kit Contains:
• 6500-2, Mainframe
• EXTL 100, Extension lead test adaptor
• SP Scan6000, Barcode scanner
• Fluke DMS 0702/PAT software
• Pass 560R, Appliance pass labels
• Fail 100S, Appliance fail labels APP 1000,
Barcode appliance number labels
(Kit contents may vary per country)
6000-2 Series PAT Testers
Recommended Accessories
Fluke 6200-2 display screen
The accuracy specification for the display range is defined as ± (%reading + digit counts) at 23 °C ± 5 °C, ≤ 75 %
RH. Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy
specification) per °C.The measurement range meets the service operating errors specified in EN61557-1: 1997,
EN61557-2: 1997, EN61557-4: 1997, EN61557-6:1997, DIN VDE0404-2.
Size (HxWxD):
200 mm x 275 mm x 100 mm
Weight:
3 kg
Two Years Warranty
See also page 62 for more details
Specifications
BDST3 Snap Tags
CableTie
PASS
Appliance 560R
Pass Labels
SP-SCAN-6000
Barcode scanner
(6500-2 Fluke only)
SP6000
Printer
Fluke DMS 0702/PAT
software
Power-on Test
The test indicates reversed L-N, missing PE, and
measures the mains voltage and frequency.
Display Range
90V to 264V
Accuracy at 50 Hz
± (2% + 3 counts)
Resolution
0.1V (1V - model 6200-2)
Input Impedance
> 1 MΩ // 2.2 μF
Maximum Input MainsVoltage 264V
Earth Bond Test
Display Range
0 to 19.99 Ω
Accuracy
(after BondTest zeroing)
± (2.5% +4 counts)
Resolution
0.01Ω
Test Current
200 mA AC - 0% + 40% into
1.99Ω
25 A AC ± 20% into 25 mΩ
at 230V
Open CircuitVoltage
> 4V AC, < 24Vac
BondTest zeroing
can subtract up to 1.99Ω
Insulation Test (Riso)
Display Range
0 to 299 MΩ
Accuracy
± (5% + 2 counts) from 0.1
to 50 MΩ
± (10% + 2 counts) from 50
to 299 MΩ
Resolution
0.01 MΩ (0 to 19.99 MΩ)
0.1 MΩ (20 to 199,9 MΩ)
1 MΩ (200 to 299 MΩ)
TestVoltage
500V DC - 0% + 10% at
500 kΩ load
250V DC - 0% + 10% at 250 kΩ
load (6500-2 only)
Test Current
> 1 mA at 500 kΩ load,
< 15 mA at 0 Ω
Auto discharge time
< 0.5 s for 1 μF
Max. Capacitive Load
Operational up to 1 μF
RCD Test: Trip Current (6500-2 only)
Operational Error
±10 %
Nominal
30 mA
Accuracy
±5 %
RCD Test: Trip Time (6500-2 only)
Standard requirement
61557 Part 6; tolerance of rated
test current 0 % to +10 %
Operational Error
±10 %
RCDType
AC General-Purpose 30 mA
Display Range
310 ms
Resolution
0.1 ms
Accuracy
3 ms
TripTimeLimitat100% (30mA)
300 ms
TripTimeLimitat500% (150 mA) 40 ms
Touch Current Test
Display Range
0 to 1.99 mA ac
Accuracy
± (4% + 2 counts)
Resolution
0.01 mA
Internal Resistance (via probe)
2 kΩ
Measuring method
Probe
The appliance under test is energized at mains potential.
Substitute Leakage Current Test
Display Range
0 to 19.99 mA ac
Accuracy
± (2.5% + 3 counts)
Resolution
0.01 mA
TestVoltage
100V AC ± 20%
Operational Error
10%
Load/Leakage Test: Load Current
Display Range
0 to 13 A*
Accuracy
± (4% + 2 counts)
Resolution
0.1 A
The appliance under test is energized at mains potential
* UK: 0 to 13 A,Austria: 0 to 10A, Germany/The Netherlands: 0 to 16A
Load/Leakage Test: Load Power
Display Range 230V mains UK:
Austria:
Germany/The Netherlands:
0.0VA to 3.2 kVA
0.0VA to 2.4 kVA
0.0VA to 3.7 kVA
Accuracy
± (5% + 3 counts)
Resolution
1VA (0 to 999VA),
0.1 kVA (>1.0 kVA)
The appliance under test is energized at mains potential
Load/Leakage Test: Leakage Current
Display Range
0 to 19.99 mA
Accuracy
± (4% + 4 counts)
Resolution
0.01 mA
The appliance under test is energized at mains potential
PELV Test
Accuracy at 50 Hz
± (2% + 3 counts)
Overload protection
300Vrms
Warning threshold
25Vrms