61

Separate hard case

The compact Fluke PAT testers are supplied with a

hard carrying case that not only offers protection

during transit but also includes extra storage space for

accessories and other tools. They’re extremely light,

weighing approximately 3 kg (without case) and have

integral carrying handles for extra convenience.

Special PAT Kit

If you need a complete PAT tester solution,

a purpose made kit is available.

Fluke 6500-2/UK Kit Contains:

• 6500-2, Mainframe

• EXTL 100, Extension lead test adaptor

• SP Scan6000, Barcode scanner

• Fluke DMS 0702/PAT software

• Pass 560R, Appliance pass labels

• Fail 100S, Appliance fail labels APP 1000,

Barcode appliance number labels

(Kit contents may vary per country)

6000-2 Series PAT Testers

Recommended Accessories

Fluke 6200-2 display screen

The accuracy specification for the display range is defined as ± (%reading + digit counts) at 23 °C ± 5 °C, ≤ 75 %

RH. Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy

specification) per °C.The measurement range meets the service operating errors specified in EN61557-1: 1997,

EN61557-2: 1997, EN61557-4: 1997, EN61557-6:1997, DIN VDE0404-2.

Size (HxWxD):

200 mm x 275 mm x 100 mm

Weight:

3 kg

Two Years Warranty

See also page 62 for more details

Specifications

BDST3 Snap Tags

CableTie

PASS

Appliance 560R

Pass Labels

SP-SCAN-6000

Barcode scanner

(6500-2 Fluke only)

SP6000

Printer

Fluke DMS 0702/PAT

software

Power-on Test

The test indicates reversed L-N, missing PE, and

measures the mains voltage and frequency.

Display Range

90V to 264V

Accuracy at 50 Hz

± (2% + 3 counts)

Resolution

0.1V (1V - model 6200-2)

Input Impedance

> 1 MΩ // 2.2 μF

Maximum Input MainsVoltage 264V

Earth Bond Test

Display Range

0 to 19.99 Ω

Accuracy

(after BondTest zeroing)

± (2.5% +4 counts)

Resolution

0.01Ω

Test Current

200 mA AC - 0% + 40% into

1.99Ω

25 A AC ± 20% into 25 mΩ

at 230V

Open CircuitVoltage

> 4V AC, < 24Vac

BondTest zeroing

can subtract up to 1.99Ω

Insulation Test (Riso)

Display Range

0 to 299 MΩ

Accuracy

± (5% + 2 counts) from 0.1

to 50 MΩ

± (10% + 2 counts) from 50

to 299 MΩ

Resolution

0.01 MΩ (0 to 19.99 MΩ)

0.1 MΩ (20 to 199,9 MΩ)

1 MΩ (200 to 299 MΩ)

TestVoltage

500V DC - 0% + 10% at

500 kΩ load

250V DC - 0% + 10% at 250 kΩ

load (6500-2 only)

Test Current

> 1 mA at 500 kΩ load,

< 15 mA at 0 Ω

Auto discharge time

< 0.5 s for 1 μF

Max. Capacitive Load

Operational up to 1 μF

RCD Test: Trip Current (6500-2 only)

Operational Error

±10 %

Nominal

30 mA

Accuracy

±5 %

RCD Test: Trip Time (6500-2 only)

Standard requirement

61557 Part 6; tolerance of rated

test current 0 % to +10 %

Operational Error

±10 %

RCDType

AC General-Purpose 30 mA

Display Range

310 ms

Resolution

0.1 ms

Accuracy

3 ms

TripTimeLimitat100% (30mA)

300 ms

TripTimeLimitat500% (150 mA) 40 ms

Touch Current Test

Display Range

0 to 1.99 mA ac

Accuracy

± (4% + 2 counts)

Resolution

0.01 mA

Internal Resistance (via probe)

2 kΩ

Measuring method

Probe

The appliance under test is energized at mains potential.

Substitute Leakage Current Test

Display Range

0 to 19.99 mA ac

Accuracy

± (2.5% + 3 counts)

Resolution

0.01 mA

TestVoltage

100V AC ± 20%

Operational Error

10%

Load/Leakage Test: Load Current

Display Range

0 to 13 A*

Accuracy

± (4% + 2 counts)

Resolution

0.1 A

The appliance under test is energized at mains potential

* UK: 0 to 13 A,Austria: 0 to 10A, Germany/The Netherlands: 0 to 16A

Load/Leakage Test: Load Power

Display Range 230V mains UK:

Austria:

Germany/The Netherlands:

0.0VA to 3.2 kVA

0.0VA to 2.4 kVA

0.0VA to 3.7 kVA

Accuracy

± (5% + 3 counts)

Resolution

1VA (0 to 999VA),

0.1 kVA (>1.0 kVA)

The appliance under test is energized at mains potential

Load/Leakage Test: Leakage Current

Display Range

0 to 19.99 mA

Accuracy

± (4% + 4 counts)

Resolution

0.01 mA

The appliance under test is energized at mains potential

PELV Test

Accuracy at 50 Hz

± (2% + 3 counts)

Overload protection

300Vrms

Warning threshold

25Vrms