Suggested test tools

Loop calibration with a

temperature transmitter at the bench

In industrial process industries, temperature

measurement equipment usually has two

components: a sensing device such as an RTD or

thermocouple and a transmitter to read and relay

the signal to the control system.

All sensors, including RTDs, drift with time.

Thus, testing the transmitter and not the sensor

could result in misjudgment regarding a system’s

performance. To avoid this potential problem,

process instrument manufacturers recommend

including the temperature sensor in loop

calibration to prove the effectiveness of the

entire system.