725/725Ex/726 Multifunction

Process Calibrators

700P27

See page 132

(Not for hazardous zones)

Specifications

725/725Ex/726 Multifunction

Process Calibrators

• Two separate channels; measure, source

and view process signals simultaneously

• Measure volts, mA, RTDs,

thermocouples, frequency, and

resistance to test sensors and

transmitters

• Source/simulate volts, mA,

thermocouples, RTDs, frequency, and

pressure to calibrate transmitters

• Measure or *source pressure using any

of 29 Fluke 700Pxx Pressure Modules

• Source mA with simultaneous pressure

measurement to conduct valve and I/P

tests

• Perform fast linearity tests with auto

step and auto ramp features

• Power transmitters during test using

24 V loop supply and simultaneous mA

measurement

• Store frequently-used test setups for

later use

• For 725Ex version see also page 134 and

135

726 Precision Multifunction

Process Calibrator

Additional features:

• More precise measurement and

calibration source performance,

accuracies of 0.01%

• Transmitter error% calculation

• Memory storage for up to 8 calibration

results

• Frequency totalizer and frequency

pulse train source mode for enhanced

flowmeter testing

• HART mode inserts 250 ohm

resistor in mA measure and source

for compatibility with HART

instrumentation

• Integrated pressure switch test allows

you to capture the set, reset and

deadband of a switch

• Custom RTD curves, add calibration

constants for certified RTD probes for

enhanced temperature measurement

Maximum voltage:

30V

Operating temperature:

10°C to 55°C

Safety:

CSA C22.2 No. 1010.1:1992

EMC: EN50082-1:1992 and EN55022:

1994 Class B

Fluke 725

Size (HxWxD):

200 mm x 96 mm x 47 mm

Weight:

0.65 kg

Battery:

Four AA alkaline batteries.

Battery life:

25 hours typical; battery door

Warranty:

Three years

Included Accessories

Test Leads, AC175 Test Clips, one pair of

stackable test leads, user’s Manuals on

CD-ROM (725Ex also includes CCD control

drawing, Statement of Quality Assurance

Practices, NIST Traceable calibration

certificate)

Ordering Information

Fluke 725 Multifunction Process Calibrator

Fluke 725Ex Intrinsically Safe Multifunction

Process Calibrator

Fluke 726 Precision Multifunction Process

Calibrator

Simultaneous Function

Capability

Channel A Channel B

24.000 mA DC

M

M or S

24.000 mA DC with

24 V loop supply

M

M or S

100.00 mV DC

M or S

30.000V DC measure

M

M or S

20.000V DC Measure

10.000V DC Source

20.000V DC Source

M or S

15 to 3200 Ohms

5 to 4000 Ohms

M or S

Thermocouple J, K,T, E, R, S, B, M,

L, U, N, XK, BP

M or S

RTD

Cu 10

, Ni120; Pt100 (392);

Pt100 (JIS); Pt100, 200, 500, 1000

(385)

M or S

Pressure (requires Fluke 700 PXX

Modules)

M M used as S

Frequency; 10 kHz;

(15 kHz)

M or S

Function

Measure or Source

Range or Type

Resolution

Accuracy

Notes

Voltage

0 to 100 mV

725: 0 to 10V (source)

0 to 20V (source)

725/726: 0 to 30V (measure)

0.01 mV

0.001 V

0.001 V

0.001 V

0.01%

0.02% Rdg + 2 LSD

Max load, 1 mA

mA

0 to 24

0.001 mA

0.01%;

0.02% Rdg

+2 LSD

Max load,

725/726: 1000Ω

725Ex: 250Ω

mV

(TC terminals)

-10.00 mV to +75.00 mV

0.01 mV

0.01%

0.02% of range +

1 LSD

Ohms

15Ω to 3200Ω

5Ω to 4000Ω

0.01Ω to 0.1Ω 0.10Ω to 1.0Ω

0.015%

Hz - CPM

2.0 to 1000 CPM

1 to 1100 Hz

1.0 to 10.0 kHz

10.0 to 15.0 kHz

0.1 CPM

1 Hz

0.1 kHz

0.1 kHz

±0.05%

±0.05%

±0.25%

±0.05%

Source;

5V p-p

1V - 20 V p-p

squarewave,

-0.1 V offset

Loop Supply

725/726: 24 V DC

725Ex: 12 V DC

N/A

10%

T/C

J, K,T, E, L, N, U, XK

0.1 ºC, 0.1 ºF

to 0.7 ºC

to 0.2 ºC

T/C

B, R, S, BP

1 ºC, 1 ºF

to 1.7 ºC

to 1.2 ºC

RTDs

Cu (10),

Ni 120 (672)

Pt 100, 200, 500, 1000 (385)

Pt 100 (3916),

Pt 100 (3926)

0.01 ºC

0.01 ºF

0.1 ºC, 0.1 ºF

to 0.15 ºC

to 0.2 ºC

Fluke 725Ex

Fluke 726

Features

M = Measure S = Source/Simulate

Unique 726 features are

in bold

725Ex: ATEX cerfified

(Ex ATEX II IG EEX 1a IIB 171°C)

Recommended Accessories

TPAK

See page 155

80PK-27

See page 150

TL220-1

See page 145

C125

See page 152

More calibration power!

*Pressure pump required

Unique 726 features are