Fluke 714B
Fluke 712B
124
712B / 714B
Temperature Calibrators
Included Accessories
Magnetic hanging tool, batteries, manual,
traceable calibration certificate, and test leads
Ordering Information
Fluke 712B RTD Temperature Calibrator
Fluke 714B TC Temperature Calibrator
Features
The new Fluke 712B and 714B are
designed for both simplicity-ease of use
and high accuracy. They are a powerful
combination of temperature and loop
calibration.
For the customer that does not want
the complexity that a multifunction test
tool brings yet demands high accuracy
for temperature calibration, these new
calibrators deliver.
Accuracy and simplicity
Recommended Accessories
700TC1
See page 151
80PK-10
See page 150
80PK-8
See page 150
The 712B can measure and simulate (13) different RTD types and resistance
The 714B can measure and simulate (17) different thermocouple types and millivolts
Measure 4 to 20 mA signals while simultaneously sourcing a temperature signal
Hanging tool designed in and included with every unit
Configurable 0 % and 100 % source settings for quick 25 % linearity checks
Linear ramp and 25 % step auto ramp based on 0 % and 100 % settings
Dual inputs and backlit display for easy interpretation of measurements
Power down settings remembered at power up for easy restart of tests
1-year and 2-year specifications and traceable certificate of calibration
Specifications
General specifications
712B
714B
Operating temperature
-10 °C to 50 °C
Operating altitude
2,000 meters
Relative humidity
(% RH operating without
condensation)
Non condensing
90 % (10 °C to 30 °C)
75 % (30 °C to 40 °C)
45 % (40 °C to 50 °C)
(Without condensation)
Vibration requirements
MIL-T-28800E, Class 2
Drop text requirements
1 meter
IP Rating
IEC 60529: IP52
Safety
IEC 61010-1, Max 30 V to earth, Pollution Degree 2
Power supply
4 AA NEDA 1.5A IEC LR6 batteries
DC mA measurement
Resolution
0-24 mA
Range
0.001 mA
Accuracy (% of reading + counts)
0.010 % + 2 μA
Ohms measurement
Ohms range
0.00 Ω to 400.00 Ω,
400.0 Ω to 4000.0 Ω
Accuracy (% of reading +counts)
0.015 % + 0.05 Ω, 0.015 % + 0.5 Ω
Ohms source
Ohms range
1.0 Ω to 400.0 Ω, 1.00 Ω to 400.00 Ω,
400.0 Ω to 1500.0 Ω,
1500.0 Ω to 4000.0 Ω
Accuracy (% of reading + counts)
0.015 % + 0.1 Ω, 0.015 % + 0.05 Ω,
0.015 % + 0.5 Ω, 0.015 % + 0.5 Ω
Resolution
0.00 Ω to 400.00 Ω
0.01 Ω
400.0 Ω to 4000.0 Ω
0.1 Ω
Millivolt measurement and source
Resolution
-10 mV to 75 mV
Range
0.01 mV
Accuracy (% of reading +counts)
0.015 % + 10 μA
New
Dimensions (HxWxD):
188.5 mm x 84 mm x 52.5 mm
Weight:
0.5 kg
Warranty:
Three years