123
724 Temperature Calibrator
80PK-3A
See page 150
Features
The Fluke 724 Temperature Calibrator is
a powerful yet easy-to-use calibrator. Use
the measure and source functions to test
and calibrate almost any temperature
instrument.
• Easy to read dual display lets you view
input and output simultaneously
• Measure RTDs, thermocouples,
ohms, and volts to test sensors and
transmitters
• Source/simulate thermocouples, RTDs,
volts, and ohms to calibrate transmitters
• Perform fast linearity tests with 25% and
100% steps
• Execute remote tests with auto step and
auto ramp
• Power transmitters during test using
loop power supply with simultaneous
mA measurement
• Store frequently-used test setups for
later use
Fluke 724
Included Accessories
Test Leads, AC175 Test Clips, one pair of
stackable test leads.
Ordering Information
Fluke 724 Temperature Calibrator
Recommended Accessories
80PK-25
See page 150
TL81A
See page 144
TL220-1
See page 145
C25
See page 152
Maximum voltage:
30V
Operating temperature:
-10°C to 55°C
Safety:
CSA C22.2 No. 1010.1:1992
EMC: EN50082-1:1992 and EN55022:
1994 Class B
Size (HxWxD):
200 mm x 96 mm x 47 mm
Weight:
0.65 kg
Battery:
Four AA alkaline batteries
Battery life:
25 hours typical; battery door
Three Year Warranty
Function
Measure or Source
Range
Resolution
Accuracy
Notes
Voltage
0 to 100 mV
0 to 10V (source)
0 to 30V (measure)
0.01 mV
0.01 V
0.001 V
0.02% Rdg + 2 LSD Max load, 1 mA
mA (measure)
0 to 24 mA
0.001 mA
0.02% Rdg +2 LSD Max load,
1000Ω
mV
-10.00 mV to +75.00 mV
0.01 mV
0.025% of range +
1 LSD
Resistance
0Ω to 3200Ω (measure)
15Ω to 3200Ω (source)
0.01Ω to 0.1Ω 0.10Ω to 1.0Ω
Loop Supply
24V DC
N/A
10%
Thermocouples
J, K,T, E, L, N, U
0.1 ºC
to 0.7 ºC
Thermocouples
B, R, S
1 ºC
to 1.4 ºC
RTDs
Ni120 (672)
Pt 100, 200, 500, 1000 (385)
Pt 100 (3916)
Pt 100 (3926)
0.1 ºC
to 0.2 ºC
Simultaneous Function
Capability
Channel A Channel B
24.000 mA DC
M
24.000 mA DC with 24V loop
supply
M
100.00 mV DC
M or S
30.000V DC Measure
M
20.000V DC Measure 10.000V
DC Source
M or S
0 to 3200 Ohms
M or S
Thermocouple J, K,T, E, R, S, B,
L, U, N
M or S
RTD Ni120; Pt100 (3926); Pt100
(JIS); Pt100, 200, 500, 1000 (385)
M or S
Specifications
M = Measure S = Source/Simulate
The temperature solution