Background Image
Previous Page  134 / 164 Next Page
Information
Show Menu
Previous Page 134 / 164 Next Page
Page Background

130

705/707/707Ex/715

Loop Calibrators

T PAK (715)

See page 155

Fluke 705

TP920

See page 144

TL220-1

See page 145

C25 (715)

See page 152

C12A (705/707)

See page 152

Functions

Voltage measurement

705/707/707 Ex

715

Range

0-28 V DC

0-200 mV, 0-20 V

Resolution

1 mV

10 μV 1 mV

Accuracy

705: 0.025% Rdg + 2 LSD

707/707Ex: 0.015% Rdg + 2 LSD

0.01% Rdg + 2 LSD

Specifications

Fluke 705, 707, 707Ex

Maximum voltage:

30 V (28 V - 707Ex)

Operating temperature:

-10 to 55°C

Safety:

CSA C22.2 No. 1010.1: 1992

EMC: EN50082-1:1992 and EN55022: 1994 Class B

Size (HxWxD):

164 mm x 75 mm x 47 mm;

Weight:

0.35 kg

Battery:

One 9 V alkaline

Battery life:

18 hours typical, at 12 mA

Three Year Warranty

Fluke 715

715

Size (HxWxD):

201 mm x 98 mm x 52 mm

Weight:

0.6 kg

Battery:

One 9V alkaline

Battery life:

4 to 20 hours

Three Year Warranty

705 Loop Calibrator

• Simultaneous mA and %

• mA accuracy of 0.02%

• Measure, source and simulate mA

• Push button 25% steps for fast, easy

linearity checks

• “Span Check” for fast confirmation of

zero and span

• Selectable slow ramp, fast ramp, and

step function

• 24 volt internal loop supply

• 0 - 20 mA or 4 - 20 mA default start up

modes

707 Loop Calibrator

• Features of the Fluke 705

• “Quick Click” detented rotary knob for

one handed operation

• Higher accuracy: 0.015%

• Mains Voltage Input protection

• 250 Ohm loop resistance for with Hart

instrumentation

715 Volt/mA Calibrator

• Measure loop current (0-20 mA,

4-20 mA) signals with 0.015% accuracy

and 0.001 mA resolution

• Measure voltage output process signals

from PLCs, transmitters

• Source or simulate 24 mA loop current

• Source voltage to 100 mV or 10 V

• 24 V loop supply with simultaneous

current measurement

707Ex Intrinsically Safe Loop

Calibrator

• Same features as Fluke 707

• For use in explosion endangered areas

• ATEX certified ( II 2 G Eex ia IIC T4)

4-20 mA, source, measure, simulate

Fluke 707

Fluke 707Ex

Current sourcing

Range

0-20 mA or 4-20 mA

1

0-20 mA or 4-20 mA

1

Accuracy

705: 0.025% Rdg + 2 LSD

707/707Ex: 0.015% Rdg + 2 LSD

0.015% + 2 LSD

Drive capability

705: 1000 Ω @ 24 mA

707: 1200 Ω @ 24 mA

707Ex: 700 Ω 20 mA

1000 Ω @ 24 mA

Loop power while measuring mA

24 V

24 V

Voltage sourcing

N/A

0-100 mV or 0-10 V

Display current and % of span

Yes

mA or %

Auto step, auto ramp

Yes

Yes

Span Check

Yes

Yes

Current measurement

Range

0-24 mA

0-24 mA

Resolution

0.001 mA

0.001 mA

Accuracy

705: 0.025% Rdg + 2 LSD

707/707Ex: 0.015% Rdg + 2 LSD

0.015%

Included Accessories

Fluke 705/707:

C10 Yellow Holster, Test Leads,

AC175 Alligator Clips, Single 9V alkaline

battery, Instruction Sheet

Fluke 707Ex:

Red Ex-holster, Test Leads, AC175

Alligator Clips, Single 9V alkaline battery,

Instruction Sheet

Fluke 715:

Yellow Holster (H80M excl. TPAK),

Test Leads and AC175 Alligator Clips, Single 9V

alkaline battery, Instruction Sheet

Ordering Information

Fluke 705

Loop Calibrator

Fluke 707

Loop Calibrator

Fluke 707Ex Intrinsically Safe Loop Calibrator

Fluke 715 Volt/mA Calibrator

1

will over-range to 24 mA

Recommended Accessories

(Not for hazardous zones)